Signal Strength
You can determine the signal strength from the symbol in the upper left corner of the display.
Troubleshooting
- Are you able to see your weight displayed on the LCD screen of the device?
- Nothing displays on the LCD when the patient steps on the scale.
- Check to ensure batteries are installed correctly.
- Worn batteries. Have the patient replace all four batteries at the same time.
- The device powers off automatically
- Low battery - have the patient replace all four batteries at the same time.
- Is the weight displayed in the range that you expect to be accurate?
- Is the device displaying in the desired unit of measurement? In the upper right hand corner of the display a patient can report the unit their device is measuring in: Pounds (lb), Stones (st lb), or Kilograms (kg).
- The unit of measurement can be toggled with the unit button located inside the battery compartment.
- Is the patient using the device on a flat, hard surface?
- Sloped surfaces or soft materials like carpet can greatly affect the accuracy of the measurements.
- After taking a measurement what code/symbol do you see on the screen?
- Error codes give insight as to what could be preventing the successful transmission of patient data - each error code signifies something different so if a patient is having technical issues with their device collecting this information can expedite the resolution process.
- Ensure that the user has tried to use the device in more than one area of the home.
- Preference is an exterior facing room near a window.
- Avoid basements or interior rooms
- Ensure the device is placed on a level and hard surface for the most accurate measurement. Soft or uneven surfaces will negatively affect the safety and accuracy of the scale.
- Have the user reboot the device to force reconnection.
- Take the batteries out of the device and press the button in the battery compartment once before putting the batteries back in.
- This will re-initiate the pairing sequence.
- Check the network status page to ensure no network outages in their area.
- If the device is continually showing E1 and the user is unable to use the device in an area with signal, start a return process.
-
- To start the return process, please send an email to the Prevounce | Pylo Support Team at support@prevounce.com .
-
LCD Display Guide and Error Codes:
Display/Error Codes | Description | Solution |
Weight overload error. The device will power off. | Retake measurement. If the issue persists, please check if your weight is out of measurement range (11 lb to 440 lb / 5 kg to 200 kg / 0 st : 11 lb to 31 st : 1 lb) |
|
SIM Error | If the issue persists, contact customer support. | |
No signal. |
1. Move the scale to an area with better signal |
|
E1 | Hardware error (such as the communication between the MCU and the module is abnormal.) |
Retake measurement. |
E3 | Data transmission error | Move to another area, preferably closer to a window. Attempt to use the device at a location where you get a strong cellular signal with your mobile phone. |
E4 | No cellular coverage or Cellular connection error |
Move to another area, preferably closer to a window. Use the device at a location where you get strong cellular signal with your mobile phone. |
E5 | The pre-loaded SIM card is replaced or the SIM service subscription suspended / terminated. |
If the issue persists, contact customer support. |
E6 | Server connection error | Move to another area, preferably closer to a window, try again. Use the device at a location where you get strong cellular signal with your mobile phone. If the issue persists, contact customer support. |
E8 | Certificate error | Contact customer support |
EErr | Hardware error | Retake measurement. If the issue persists, contact customer support |